HT _ 110 TOC ANALYZER

brand

HT-110 — Specifications

Specification Exact Details
Model HT-110
Measurement method Continuous UV-Oxidation and measurement of conductivity differential, according to USP 643, EP 2.2.44, JP16
Measuring range 0 to 1000 ppbC(Measuring range : 0 to 100/500/1000 ppbC)
Display resolution 0 to 100 ppbC — 0.1 ppbC0 to 500 ppbC — 0.1 ppbC0 to 1000 ppbC — 1 ppbC
Repeatability Within ±2% of the full scale
Analog output Number of outputs : 14 to 20mA DC / 0 to 1V DC (option) : Input/Output isolated typeMax. load resistance : (4 to 20mA DC) 600Ω. Min. load resistance (0 to 1V DC) : 100kΩ [with option selected]Analog output range : Freely selectable within the measurement rangeOutput contents : TOC concentration
Contact output Number of outputs : 2Alarm contact output (RLY1)Contact type : Relay contact, SPDTContact rating : 125V AC 3A, 24V DC 3A (resistance load)Contact function : Upper/lower limit action (ON/OFF control)Output contents : Selectable between TOC concentration, Temperature, Conductivity, Hold, and Error(Operation is reversed when Hold or Error is selected)Alarm contact output (RLY2)Contact type : Relay contact, SPSTContact rating : 125V AC 3A, 24V DC 3A (resistance load)Contact function : Upper/lower limit action (ON/OFF control)Output contents : TOC concentration
Contact Input Number of input point : 1
Sample conditions Conductivity : < 10 µS/cm or lower (0 to 100 ppb: < 2 µS/cm or lower)Temperature : 5 to 99 °CPressure : 0.03 MPa to 0.50 MPaSample flow rate : 10 to 30mL/min (Recommended 15mL/min for measurement)
Ambient environment Temperature : 5 to 40°CHumidity : 85% or lower
Display functions LC with backlightDisplay content (TOC concentration, conductivity, temperature, alarm)
Structure Protection : IP43 ; Paint color 5PB8/1; Epoxy coating (indoor-use panel installation type)
Power supply 100 to 240V AC ±10% 50/60 Hz 80 VA (max.)
Mass 13 kg
Conforming standards CE marking, FCC Part 15